Medical Device Fundamentals
MDR Technical Documentation: Structure, Content and Key Requirements
What Annexes II and III of Regulation (EU) 2017/745 contain, how to link the documents to one another and how to keep the file up to date throughout the lifetime of the device.
The technical documentation is the set of documents through which the manufacturer demonstrates that its device meets the requirements of Regulation (EU) 2017/745. The notified body assesses it depending on the class of the device, the competent authorities may request it, and the manufacturer keeps it for at least ten years after the last device covered has been placed on the market, or fifteen years for implantable devices (Article 10(8)).
Its content is set out in Annexes II and III. The Regulation does not impose a format, but requires the documentation to be presented in a clear, organised, readily searchable and unambiguous manner. This article describes its structure, what each part contains and how to keep it up to date; the most common gaps are covered in a separate article.
What the technical documentation is for
It serves three purposes. It demonstrates conformity with the general safety and performance requirements (GSPR) of Annex I. It provides the basis for the conformity assessment, which the notified body carries out depending on the class of the device. And it enables the manufacturer to keep control of its product over time: every change and every piece of surveillance data is reflected in it.
Under the former Directives, the term “technical file” was widely used; the MDR consistently speaks of technical documentation. Exchanges with notified bodies rely on a set of established shorthand terms, such as Annex II, GSPR, PMS plan and PSUR, which this article uses as well.
A free format, with useful reference points
The Regulation does not prescribe an outline. Following the order of Annexes II and III is the most direct solution; the IMDRF table of contents for regulatory submissions (2024 edition) is useful when the same file serves several markets. The notified bodies grouped in Team-NB also publish best practice guidance on the presentation of technical documentation (version 4, April 2026).
In practice, one set of technical documentation most often covers the devices grouped under a single Basic UDI-DI, with their variants and configurations, each of which is described and covered by the evidence.
A five-layer reading grid
To check whether a file holds together, it helps to read it not section by section, but as a chain of demonstration. The grid below groups the sections of Annexes II and III into five layers that build on one another.
Reading the technical documentation in five layers
Traceability: intended purpose → requirement → risk control → evidence → safety information
- ReferenceAnnex II, Section 1
- Intended purpose and indications
- Description, variants and accessories
- Class and rule applied
- Basic UDI-DI
- RequirementsAnnex II, Section 4
- Applicable GSPR and justifications
- Harmonised standards and other solutions
- Common specifications
- EvidenceAnnex II, Sections 3, 5 and 6
- Design and manufacture
- Benefit-risk and risk management
- Verification and validation
- Clinical evaluation
- InformationAnnex II, Section 2
- Labels and packaging
- Instructions for use
- UDI carriers
- LifecycleAnnex III
- PMS plan and PMCF
- PMS report or PSUR
- Changes and updates
Field data feed back into the first four layers.
Traceability runs vertically. A claim in the instructions for use must be found in the intended purpose, be covered by the clinical evaluation and rest on verification evidence; a risk control measure must appear in the design requirements, be verified and, if it takes the form of a warning, appear in the instructions for use. When one of these links is missing, the file loses its demonstrative force, even if each document seems complete.
The content, block by block
The following table lists each documentation block, its regulatory basis, its typical content and the typical gaps:
| Block | Regulatory basis | Typical content | Typical gaps |
|---|---|---|---|
| Description and specification | Annex II, Section 1 | Intended purpose, patient population, users, principle of operation, justification of the qualification and class, variants, accessories, key components, materials in contact with the body, specifications, previous generations | Intended purpose worded differently across documents; variants not covered by the evidence |
| Information supplied by the manufacturer | Annex II, Section 2; Annex I, Chapter III | Labels on the device and its packaging, instructions for use in the required languages | Safety information with no link to the risk management file |
| Design and manufacturing | Annex II, Section 3 | Design stages, manufacturing processes and their validation, final inspection, sites, suppliers and subcontractors | Special processes without documented validation; sites or subcontractors not identified |
| GSPR | Annex II, Section 4; Annex I | Applicability of each requirement, method of demonstration, standards or specifications applied, references of the evidence documents | Requirements declared as met without reference to an identified document; non-applicability not justified |
| Benefit-risk and risk management | Annex II, Section 5; Annex I, Sections 1 to 9 | Risk management plan, risk analysis, evaluation and control, residual risks, benefit-risk conclusion | Control measures without evidence of effectiveness; production and post-market risks missing |
| Pre-clinical verification and validation | Annex II, Section 6.1(a) and (b); Section 6.2 | Engineering and laboratory tests, simulated use, material characterization, software, information for specific cases | Reports without a protocol or acceptance criteria; device version not identified |
| Clinical evaluation and PMCF | Annex II, Section 6.1(c) and (d); Annex XIV | Clinical evaluation plan and report and their updates; PMCF plan and PMCF evaluation report, or a justification of why PMCF is not applicable | Equivalence insufficiently demonstrated; report out of step with the intended purpose and claims |
| Post-market surveillance | Annex III | PMS plan (sources, indicators, thresholds, methods), PMS report or PSUR | Generic plan; results with no effect on risk management or the clinical evaluation |
The device description, the anchor point
Section 1 of Annex II is not a sales presentation: it is the reference definition of the device. Three elements need particular care: an intended purpose worded once and for all, with the patient population, users and use environment; a written justification of the qualification and of the classification rule applied; and a description of variants and configurations precise enough for every piece of evidence to be linked to it. If this section is imprecise, no evidence can be linked unambiguously to the device.
The GSPR checklist
For each requirement of Annex I, Section 4 of Annex II states whether it applies and why, the method used to demonstrate conformity, the harmonised standards, common specifications or other solutions applied, and the precise identity of the controlled documents that provide the evidence. Where the standard applied is not harmonised, or a more recent edition exists, the checklist justifies the choice against the state of the art. It is the navigation table of the file: a reviewer should be able to start from a requirement and reach the evidence in a few steps.
Verification, validation and clinical data
Section 6 brings together the results of engineering, laboratory, simulated use and animal tests, each with its protocol, methods of analysis and conclusions, together with software verification and validation. Where no new testing has been undertaken, the file explains why. Section 6.2 adds information for specific cases: medicinal substances, tissues or cells of human or animal origin, substances absorbed by the body or locally dispersed, the presence of substances that are carcinogenic, mutagenic, toxic to reproduction or endocrine-disrupting, sterile devices, devices with a measuring function, and devices to be connected to other devices.
Clinical data are found in the clinical evaluation report, its updates and its plan, which are always required, and in the post-market clinical follow-up (PMCF) plan and PMCF evaluation report, whose absence must be justified. Each test report refers to an identified version of the device and to the requirement it demonstrates.
Annex III: the post-market surveillance documentation
The technical documentation does not stop at placing on the market. Annex III adds the post-market surveillance plan and, depending on the class, the PMS report (class I) or the PSUR (classes IIa, IIb and III). What matters in the file is the loop: each conclusion of the PMS report or PSUR is reflected in risk management, the clinical evaluation or the documentation, or is explicitly assessed as having no effect on them.
Software, usability and cybersecurity
Three topics call for dedicated files, integrated into the technical documentation and linked to the risk management file.
Software
Annex I, Section 17, requires software to be developed and manufactured in accordance with the state of the art, taking into account the principles of the development lifecycle, risk management, including information security, verification and validation. IEC 62304 generally serves as the reference: development plan, requirements, architecture, verification, management of third-party software, release and maintenance. The qualification and classification of software follow MDCG 2019-11 rev. 1.
Usability
The general requirements aim to reduce the risks related to ergonomic features and the environment of use, as well as use errors. The usability engineering file, generally prepared in accordance with IEC 62366-1, describes the use specification, the user interface, the analysis of use-related risks and the evaluations carried out during design.
Cybersecurity
For devices that incorporate electronic programmable systems and for software that is a device in itself, Annex I requires information security to be taken into account in development (Section 17.2), minimum requirements concerning hardware, IT network characteristics and IT security measures, including protection against unauthorized access, to be set out (Section 17.4), active devices to be protected as far as possible against unauthorized access (Section 18.8) and users to be informed in the instructions for use (Section 23.4). The file presents the security risk analysis, the requirements and measures adopted, their verification and the information intended for users. MDCG 2019-16 rev. 1 sets out the expectations; IEC 81001-5-1 serves as the reference for security activities in the software lifecycle.
Keeping the documentation up to date
Technical documentation is never static. It must be possible to reconstruct each version: which version of the file corresponded to which version of the device, and on what evidence it relied. The triggers for an update are well known:
- Conclusions from post-market surveillance, PMCF and PSURs.
- Corrective and preventive actions affecting design, manufacturing or labeling.
- Changes to design, suppliers, processes or software.
- New or revised standards, which move the state of the art forward.
- Regulatory developments: amending acts, common specifications, MDCG documents.
- Changes affecting UDI, which may require a new UDI-DI.
Where a notified body has been involved, changes are brought to its attention as agreed with it; for devices covered by an EU technical documentation assessment certificate, changes that could affect safety, performance or the conditions prescribed for use require its prior approval (Annex IX, Section 4.10).
Preparing for the notified body review
The notified body first checks that the file is complete, then reviews it in substance and raises questions, often over several rounds. For classes IIa and IIb, it assesses representative devices selected according to a sampling plan; for class III and class IIb implantable devices, it assesses every device, except in the cases provided for in Article 52(4). For certain class III and class IIb devices, the clinical evaluation is also subject to consultation of an expert panel (Article 54); for implantable and class III devices, the draft summary of safety and clinical performance (Article 32) is part of the submitted file.
Three elements make this review easier: a general overview of the device, its variants and the organization of the file; precise cross-references, from the GSPR checklist to the evidence and from one document to another; and consistent, identified versions throughout the file. The Team-NB guidance on the presentation of technical documentation is a useful benchmark for this preparation.
Self-check questions
Before submitting or updating a file, these questions help to check how robust it is:
- Is the intended purpose worded once, then reused word for word everywhere?
- Is every variant and configuration described and covered by the evidence?
- Does each applicable requirement of Annex I refer to a controlled, identified document?
- Is every non-applicability justified?
- Is every risk control measure verified, and its effectiveness demonstrated?
- Does every test report refer to a device version and to a protocol with acceptance criteria?
- Does the clinical evaluation cover exactly the intended purpose, indications and claims?
- Does the PMS plan define indicators, thresholds and the decisions they trigger?
- Can you reconstruct which version of the file corresponded to which version of the device?
Regulatory references
- Regulation (EU) 2017/745, Annexes I, II and III (consolidated version of 9 November 2026)eur-lex.europa.eu
- MDCG guidance documents, including MDCG 2019-11 rev. 1 and MDCG 2019-16 rev. 1health.ec.europa.eu
- Team-NB, Best Practice Guidance on technical documentation, version 4 (April 2026)team-nb.org
- IMDRF, table of contents for regulatory submissions (N9, edition 4)imdrf.org
- ISO 14971:2019, application of risk management to medical devices (ISO)iso.org