EU MDR 2017/745: Understanding the Key Requirements for Medical Devices

What the European Medical Device Regulation requires of manufacturers, from the intended purpose of a product to post-market surveillance, and the order in which its requirements build on one another.

Published

Regulation (EU) 2017/745, the Medical Device Regulation (MDR), governs the placing on the market, making available and putting into service of medical devices in the European Union. It has applied since 26 May 2021 and replaced Directives 93/42/EEC and 90/385/EEC. Like any regulation, it applies directly in every Member State; national legislation only supplements it where the MDR leaves room to do so, for example on the languages of the information supplied with the device or the organization of the competent authorities.

This article does not comment on the text article by article. It follows the order in which questions arise for a manufacturer: qualifying the product, determining its class, demonstrating conformity, placing it on the market, then monitoring it and managing its changes. Article and annex numbers are cited only where they help to locate a requirement.

What the MDR regulates

The Regulation pursues two inseparable objectives: a high level of protection of the health of patients and users, and the smooth functioning of the internal market. Compared with the Directives, it strengthens the requirements for clinical evidence, post-market surveillance, device traceability and transparency of information, and it regulates the designation and monitoring of notified bodies more closely.

Above all, it takes a lifecycle approach: the manufacturer’s obligations begin with design and continue for as long as the device is on the market.

The MDR across the lifecycle of a device

  1. Qualification and classArt. 2 · Art. 51 · Annex VIIIIntended purpose, medical device status, applicable classification rule.
  2. Design under a quality systemArt. 10 · Annex IGeneral safety and performance requirements, risk management, verification and validation.
  3. Clinical evaluationArt. 61 · Annex XIVSufficient clinical data, clinical benefit, benefit-risk ratio.
  4. Conformity assessmentArt. 52 · Annexes IX to XITechnical documentation and, depending on the class, involvement of a notified body.
  5. Placing on the marketArt. 19 · 20 · 27 · 29EU declaration of conformity, CE marking, UDI, registration in EUDAMED.
  6. Post-market surveillance and vigilanceArt. 83 to 92PMS plan, PMCF, periodic reports, serious incidents, trends.
  7. ChangesAnnex IX, Sections 2.4 and 4.10Impact assessment, information to the notified body, update of the documentation.

Feedback: field data and changes update risk management, the clinical evaluation and the technical documentation.

Seven stages, from qualifying the product to managing its changes. Field data and changes lead back to the design and demonstration stages: the loop stays open for as long as the device is on the market.

Scope in brief

A product falls under the MDR when its manufacturer intends it for one or more of the specific medical purposes defined in Article 2, such as the diagnosis, prevention, monitoring, prediction, prognosis, treatment or alleviation of disease, or the diagnosis, monitoring, treatment, alleviation of, or compensation for, an injury or disability, and when its principal intended action in or on the human body is not achieved by pharmacological, immunological or metabolic means. Accessories for medical devices and certain products without an intended medical purpose, listed in Annex XVI, also fall within its scope.

Software can be a medical device in its own right; its qualification and classification follow the same logic, detailed in MDCG 2019-11 rev. 1 (June 2025). The Regulation applies in the Union and in the other states of the European Economic Area; other countries have their own framework, such as Switzerland with its Medical Devices Ordinance (MedDO).

The starting point: intended purpose, qualification and class

Everything starts with the intended purpose: the use for which a device is intended according to the data supplied by the manufacturer on the label, in the instructions for use or in promotional or sales materials or statements, and as specified in the clinical evaluation. It determines whether the product is a medical device, which classification rule applies and which clinical data will be needed. An intended purpose that is vague, or worded differently from one document to another, weakens everything that follows from it.

Classification is based on the 22 rules of Annex VIII. They take into account the duration of use, invasiveness, the part of the body concerned, whether the device is active and certain particular characteristics. The result is one of four classes, I, IIa, IIb or III; where several rules apply, the strictest one prevails. MDCG 2021-24, revised in April 2026, illustrates how each rule is applied.

The class determines more than whether a notified body is involved. It also sets the depth of the technical documentation assessment, the form of the post-market surveillance reports and the obligation to publish a summary of safety and clinical performance:

What varies with the risk class (simplified view)
RequirementClass IClass IIaClass IIbClass III
Notified body involvementNo, except for aspects relating to sterility, metrology or the reuse of surgical instrumentsYesYesYes
Notified body assessment of the technical documentationLimited to the aspects aboveOn a sampling basis: at least one representative device per category of devicesOn a sampling basis per generic device group; every implantable device, with exceptionsEvery device
Post-market surveillance reportingPMS report, updated when necessaryPSUR, at least every two yearsPSUR, at least annuallyPSUR, at least annually
Summary of safety and clinical performance (SSCP)NoFor implantable devicesFor implantable devicesYes

This view is deliberately simplified: custom-made devices, systems and procedure packs, devices incorporating a medicinal substance and Annex XVI products follow specific rules.

An overview of the requirements

The general obligations of manufacturers are set out in Article 10. They are easier to read by area, asking each time what the requirement is for and what evidence it produces:

MDR requirement areas, purpose and typical evidence
AreaPurposeTypical evidence
Quality management system (Article 10(9))Control design, manufacture, surveillance and changes, in proportion to the risk class and the type of deviceQuality manual, procedures, records; certificate from a notified body where the class requires it
Risk management (Annex I, Sections 1 to 9)Reduce risks as far as possible without adversely affecting the benefit-risk ratioRisk management plan, analysis, control measures and report, usually in accordance with ISO 14971
General safety and performance requirements (Annex I)Demonstrate that the device is safe and achieves its intended performanceGSPR checklist referring to the standards applied and the evidence documents
Clinical evaluation (Article 61, Annex XIV)Establish safety, performance and clinical benefit on the basis of sufficient dataClinical evaluation plan and report, PMCF plan and PMCF evaluation report
Technical documentation (Annexes II and III)Bring the demonstration of conformity together in a clear, searchable formStructured file, kept up to date, available to the authorities and the notified body
Identification and registration (Articles 27 to 31, Annex VI)Ensure traceability and transparencyBasic UDI-DI, UDI-DI and UDI-PI, single registration number (SRN), registration data in EUDAMED
Information supplied with the device (Annex I, Chapter III; Article 18)Enable safe use and identify the manufacturerLabels, instructions for use, implant card for the implantable devices concerned
Post-market surveillance (Articles 83 to 86)Actively collect and analyze field data throughout the lifetime of the devicePMS plan, PMCF, PMS report or PSUR
Vigilance (Articles 87 to 92)Report and handle serious incidents and field safety corrective actionsVigilance procedure, incident reports, trend reports, field safety notices
Person responsible for regulatory compliance (Article 15)Ensure that a qualified person oversees the key aspects of complianceDesignation, evidence of qualification, defined responsibilities
Financial coverage (Article 10(16))Cover the manufacturer’s liability for damage caused by a defective deviceInsurance or other measures proportionate to the risk class, the type of device and the size of the enterprise

These pieces of evidence do not simply sit side by side. Risk management feeds the design requirements and the warnings in the instructions for use; the clinical evaluation rests on the same claims as the labeling; post-market surveillance updates all three. It is this overall consistency that notified bodies and authorities examine.

Demonstrating conformity: GSPR, risks and clinical evaluation

Annex I sets out the general safety and performance requirements (GSPR). Chapter I lays down the general requirements (Sections 1 to 9), including risk management. Chapter II covers design and manufacture (Sections 10 to 22): chemical, physical and biological properties, infection and microbial contamination, electronic programmable systems and software, and protection against mechanical and thermal risks, among others. Chapter III covers the label and the instructions for use (Section 23).

Harmonised standards, common specifications and the state of the art

Applying a harmonised standard whose reference has been published in the Official Journal gives a presumption of conformity with the requirements it covers (Article 8), as specified in its Annex Z; this is the case for EN ISO 13485 for the quality management system. Common specifications adopted by the Commission (Article 9) apply where they exist, unless the manufacturer duly justifies solutions that ensure a level of safety and performance at least equivalent to them. For everything else, the manufacturer demonstrates conformity against the state of the art, often using international standards that are not harmonised under the MDR, such as IEC 62304 for the software lifecycle or IEC 62366-1 for usability.

Risk management, the common thread of the file

The MDR requires a risk management system that is established, implemented, documented and maintained throughout the lifetime of the device (Annex I, Section 3). Risks are reduced as far as possible without adversely affecting the benefit-risk ratio, in order of priority: safety by design, then protective measures, then information for safety. Production and post-market surveillance data are continuously fed back into it.

ISO 14971:2019 is the de facto reference; its European version, EN ISO 14971:2019 + A11:2021, is harmonised, and its Annex Z sets out how it relates to the requirements of the Regulation.

Clinical evaluation, a continuous process

Clinical evaluation (Article 61, Annex XIV, Part A) demonstrates, on the basis of sufficient clinical data, that the device achieves its performance, that it is safe and that its clinical benefit justifies the residual risks. It follows a plan, draws on the literature, on clinical investigations and on post-market clinical follow-up (PMCF) data, and results in a report that is updated throughout the lifetime of the device.

For implantable and class III devices, clinical investigations are required in principle, except in the cases provided for in Article 61; Delegated Regulation (EU) 2026/1451 extended the list of exempted devices to certain well-established technologies. Claiming equivalence with another manufacturer’s device requires demonstrating equivalent technical, biological and clinical characteristics; for these devices, there must also be a contract giving full and ongoing access to the other manufacturer’s technical documentation, and the original clinical evaluation must have been performed in compliance with the Regulation.

Conformity assessment and the role of the notified body

Before affixing the CE marking, the manufacturer follows the conformity assessment procedure laid down for the class of its device (Article 52). For a class I device that is not sterile, has no measuring function and is not a reusable surgical instrument, the manufacturer draws up the technical documentation and then the EU declaration of conformity under its sole responsibility. In all other cases a notified body is involved, most often under Annex IX, which combines the assessment of the quality management system with the assessment of the technical documentation; Annexes X and XI provide for type-examination and product conformity verification.

The notified body must be designated for the codes that correspond to the device, which can be checked in the Commission’s NANDO database. Its certificates are valid for no more than five years (Article 56); it carries out surveillance audits at least once every twelve months and unannounced audits at least once every five years (Annex IX, Section 3). At the end of the procedure, the manufacturer draws up the EU declaration of conformity (Article 19, Annex IV) and affixes the CE marking (Article 20). It remains responsible for the conformity of its device.

Placing on the market: UDI, EUDAMED and economic operators

Placing a device on the market involves more than CE marking. The device must be identifiable and registered, and every actor in the supply chain must know its obligations.

UDI: identifying each device

The Unique Device Identification system (Article 27, Annex VI) combines several identifiers. The Basic UDI-DI, the main identifier of a device model, appears in the declaration of conformity, the certificates and the technical documentation; according to MDCG 2018-1, it groups devices that share the same intended purpose, risk class and essential design and manufacturing characteristics. The UDI-DI is specific to a device and its packaging level; the UDI-PI identifies the unit of production, for example the lot or batch, the serial number or the date of manufacture.

UDI carriers appear on the label and on the levels of packaging. They have been required for all classes since 26 May 2025; direct marking of reusable class I devices applies from 26 May 2027.

EUDAMED: a module-by-module roll-out

The European database EUDAMED (Article 33) has been rolled out gradually since Regulation (EU) 2024/1860. In Decision (EU) 2025/2371, published in the Official Journal on 27 November 2025, the Commission confirmed that four modules are functional: actor registration, UDI and device registration, notified bodies and certificates, and market surveillance. Their use has been mandatory since 28 May 2026.

Devices placed on the market before that date that continue to be placed on the market must be registered no later than twelve months after publication of the notice, i.e. by 28 November 2026; notified bodies have eighteen months, until 28 May 2027, to enter the certificates they issued earlier. The vigilance and post-market surveillance module and the clinical investigations module are still in preparation; until they are available, the corresponding exchanges of information, notably incident reporting, follow the provisions of the former Directives (Article 123(3)(d)).

Economic operators and the PRRC

The MDR divides obligations between the manufacturer, the authorised representative, the importer and the distributor (Articles 10 to 14). A manufacturer established outside the Union designates a single authorised representative established in the Union, which in particular verifies the declaration of conformity and the technical documentation, keeps these documents available to the authorities and can be held liable for defective devices, jointly and severally with the manufacturer (Article 11). Before placing a device on the market, the importer verifies the CE marking, the declaration of conformity, the labeling, the instructions for use, the UDI and the registration (Article 13); the distributor carries out proportionate verifications (Article 14). A distributor, importer or other person takes on the manufacturer’s obligations when it makes a device available under its own name or trademark, changes its intended purpose or modifies it in a way that may affect its compliance (Article 16).

Each manufacturer has at least one person responsible for regulatory compliance (PRRC), whose qualifications are set out in Article 15; micro and small enterprises may rely on an external person who is permanently and continuously at their disposal. Since 10 January 2025, Article 10a also requires the manufacturer to inform the competent authority and the economic operators, health institutions and healthcare professionals to whom it directly supplies the device, as a rule at least six months in advance, of an interruption or discontinuation of supply where it is reasonably foreseeable that this could result in serious harm or a risk of serious harm to patients or public health.

After placing on the market: post-market surveillance, PMCF and vigilance

Post-market surveillance is a system that the manufacturer plans, establishes, documents, implements, maintains and updates in a manner proportionate to the risk class and appropriate for the type of device, as part of its quality management system (Article 83). It actively gathers data on quality, performance and safety, analyzes them and draws conclusions from them: updates to the benefit-risk determination, risk management, the clinical evaluation and the documentation, and corrective and preventive actions.

The PMS plan (Annex III) sets out the data sources, methods, indicators and trigger thresholds. PMCF, which is part of it, gathers clinical data after the device has been placed on the market. The results are summarized in a post-market surveillance report for class I and in a periodic safety update report (PSUR) for classes IIa, IIb and III. MDCG 2025-10 (December 2025) sets out the expectations in this area.

Vigilance

The manufacturer reports every serious incident to the competent authorities, except expected side-effects that are subject to trend reporting (Article 87): immediately after establishing a causal relationship, or the reasonable possibility of one, and no later than fifteen days after becoming aware of the incident; no later than two days in the event of a serious public health threat; and no later than ten days in the event of death or an unanticipated serious deterioration in a person’s state of health. Field safety corrective actions are reported before they are undertaken, except in cases of urgency. The manufacturer also reports any statistically significant increase in the frequency or severity of incidents that are not serious incidents or that are expected undesirable side-effects (Article 88), and informs users by means of a field safety notice when a field safety corrective action is taken (Article 89).

Changes over the lifecycle

A device evolves: suppliers, processes, software, intended purpose, labeling. Every change goes through the change control of the quality system, which assesses its impact on conformity with the GSPR, risk management, the clinical evaluation, the technical documentation and UDI; some changes require a new UDI-DI to be assigned.

Where a notified body has been involved, the manufacturer informs it of any plan for substantial changes to the quality management system or to the range of devices covered (Annex IX, Section 2.4). For devices covered by an EU technical documentation assessment certificate, i.e. class III devices and non-exempted class IIb implantable devices, changes that could affect safety, performance or the conditions prescribed for use also require the notified body’s approval (Annex IX, Section 4.10). For devices still under the transitional provisions, a significant change in design or intended purpose ends that regime; MDCG 2020-3 rev. 1 helps to determine whether a change is significant.

Where the regulatory framework stands

The text of the Regulation has been amended several times since 2017. As of 11 November 2026, the following points are worth knowing:

  • “Legacy” devices: devices covered by a certificate issued under the former Directives, and devices whose declaration of conformity was drawn up before 26 May 2021 without a notified body but which require one under the MDR, may be placed on the market until 31 December 2027 (class III and most class IIb implantable devices) or until 31 December 2028 (other devices concerned). The conditions include: continuing to comply with the Directive, with no significant change in design or intended purpose; having put in place, by 26 May 2024, a quality management system in accordance with Article 10(9) and lodged an application by that same date; and having signed a written agreement with a notified body by 26 September 2024. The MDR requirements on post-market surveillance, vigilance and registration apply to these devices (Regulation (EU) 2023/607).
  • EUDAMED: four modules mandatory since 28 May 2026; devices already on the market to be registered by 28 November 2026.
  • Article 10a: prior notification of an interruption or discontinuation of supply, applicable since 10 January 2025.
  • Electronic instructions for use: Implementing Regulation (EU) 2025/1234 extends them to all devices intended for professional users.
  • Notified bodies: Implementing Regulation (EU) 2026/977 harmonises certain of their procedural requirements from 25 February 2027.
  • Well-established technologies: Delegated Regulations (EU) 2026/1359 and 2026/1451 extend the list of class IIb implantable devices exempted from the assessment of the technical documentation of every device, and the list of devices exempted from clinical investigations.
  • Substances: since 1 January 2026, Regulation (EU) 2025/2457 has amended Annex I, Section 10.4, on substances that are carcinogenic, mutagenic or toxic to reproduction and on endocrine-disrupting substances.
  • Targeted revision: the Commission proposal of 16 December 2025 (COM(2025) 1023) is being examined by the European Parliament and the Council; until it has been adopted and published, it does not change the applicable law.

What is often underestimated

Some points come up regularly when the Regulation is implemented:

  • Class I: in most cases no notified body is involved, but the quality management system, technical documentation, clinical evaluation and post-market surveillance are still required.
  • Languages: each Member State determines the languages of the labeling and instructions for use for devices made available on its territory (Article 10(11)).
  • Clinical data for “legacy” devices, which are often insufficient against current requirements.
  • Post-market surveillance reduced to a complaints log instead of active, planned data collection.
  • Critical suppliers and subcontractors, which the notified body may audit, including through unannounced audits.
  • The quality of the data entered in EUDAMED, much of which becomes public.
  • The duration of the notified body review and its rounds of questions, which need to be built into the project schedule.

Practical questions for a manufacturer

These questions help to position a project and set priorities:

  • Where to start? With a precisely worded intended purpose and a written justification of the qualification and the class.
  • Which documents and processes must work together? The quality management system, risk management, clinical evaluation, post-market surveillance and technical documentation, with explicit cross-references between them.
  • Which obligations continue after placing on the market? Post-market surveillance, PMCF, periodic reports, vigilance, registrations in EUDAMED, change control and updating the documentation.
  • Where do quality, risk management, clinical evaluation and post-market surveillance meet? In management review, in CAPA and in every update of the clinical evaluation report and the risk management file.
  • When should specialist support be considered? Before a first notified body assessment, when the class or intended purpose changes, when nonconformities have to be addressed quickly, or when the team lacks capacity for a specific deadline.